XSelect CSH XP 2.5 µm Columns

3

[ CARE AND USE MANUAL ]

System Volume

4.4 %h

5

Figure 2: Determination of system band spread volume using 5-Sigma Method.

Table 1: Expected System Band Spread Volumes

System

Band Spread

Alliance 2695 HPLC

29 µL

Vendor A HPLC

41 µL

Vendor B UHPLC (600 bar)

28 µL

Vendor C UHPLC

21 µL

Vendor D UHPLC

17 µL

ACQUITY UPLC

12 µL

ACQUITY UPLC H-Class

9 µL

ACQUITY UPLC I-Class (FTN)

7.5 µL

ACQUITY UPLC I-Class (FL)

5.5 µL

e. Measuring System Dwell Volume

Dwell volume is different than system band spreading. System dwell
volume is a measurement of the volume it takes for the initial gradient
conditions to reach the head of the column. This calculation is
particularly useful when it is necessary to transfer a method between
different LC systems.

1. Disconnect the column from the system and replace with a zero dead

volume union.

2. Use acetonitrile as mobile phase A, and acetonitrile with 0.05 mg/mL

uracil as mobile-phase B.

3. Monitor UV at 254 nm.

4. Use the flow rate in the original method and the intended flow rate on

the target instrument.

5. Collect 100% A baseline for 5 minutes.

6. At 5 minutes, program a step to 100% B, and collect data for an

additional 5 minutes.

7. Measure absorbance difference between 100% A and 100% B.

8. Measure time at 50% of that absorbance difference.

9. Calculate time difference between start of step and 50% point.

10. Multiply time difference by flow rate to calculate system volume.

Programmed time = 5 minutes

50% Absorbance = 0.35852 AU
Time = 5.6953 minutes

100% Asymptotic

Total absorbance = 0.7164 AU

5.69
5.00
0.69 min

= Programmed Gradient

= Actual Gradient

System Volume
0.69 min x 1.5 mL/min = 1.04 mL

0.70

0.65

0.60

0.55

0.50

0.45

0.40

0.35

0.30

0.25

0.20

0.15

0.10

0.05

0.00

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

18.00

20.00

Figure 3: Measuring system band spread volume.

Table 2: Expected System Dwell Volumes

System

Dwell Volume

Alliance 2695 HPLC

900 µL

ACQUITY UPLC

120 µL

ACQUITY UPLC H-Class

350 µL

ACQUITY UPLC I-Class (FTN)

100 µL

ACQUITY UPLC I-Class (FL)

95 µL

f. Column Equilibration

XSelect CSH XP 2.5 µm Columns are shipped in 100% acetonitrile.
It is important to ensure mobile-phase compatibility before changing
to a different mobile-phase system. Equilibrate the column with a
minimum of 10 column volumes of the mobile phase to be used
(refer to Table 3 for a list of column volumes). The column may
be considered fully equilibrated once a constant backpressure
is achieved.