beautypg.com

Care and use manual – Waters Oligonucleotide Separation Technology XBridge OST C18 Columns User Manual

Page 7

background image

[ CARE AND USE MANUAL ]

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com

Waters is a registered trademark of Waters Corporation. Empower, BioAlliance,
XBridge and The Science of What’s Possible are trademarks of Waters
Corporation. Acrodisc is a trademark of Pall Life Sciences. SLIPFREE is a
trademark of Keystone, Inc. PEEK is a trademark of Victrex plc.

©2012 Waters Corporation. Printed in the U.S.A.
June 2012 720001463 SC-IH

c. Impact of Bandspreading Volume on 2.1 mm i.d.
Column Performance

Use of a 2.1 mm i.d. XBridge OST C

18

column may require modification

to the HPLC system tubing in order to eliminate excessive system
bandspreading volume. Without proper system modification,
excessive system band-spreading volume will cause peak broadening
and will have a detrimental impact on peak resolution. For example,
an HPLC System with 70 μL bandspreading might yield 10,000
plates while the same column tested on a LC System with 130 μL
bandspreading would yield only 8,000 plates.


Note: Flow splitters after the column will introduce additional bandspreading.
System optimization, especially in a system that contains a flow splitter,
can have dramatic effects on sensitivity and resolution. Optimization includes
using correct ferrule depths and minimizing tubing inner diameters and
lengths. An example is given in Figure 9 where system optimization
resulted in a doubling of component sensitivity and resolution in an LC/
MS/MS system.


Figure 9: Non-Optimized vs. Optimized LC/MS/MS System

Table 3: XBridge OST C

18

Ordering Information

Description

Particle Size Pore Size Dimension Part No.

XBridge OST C

18

2.5 μm

135Å

2.1 x 50 mm

186003952

XBridge OST C

18

2.5 μm

135Å

4.6 x 50 mm

186003953

XBridge OST C

18

2.5 μm

135Å

10.0 x 50 mm 186003954

Custom
XBridge OST C

18

--

--

--

186003955

7.0

7.5

Non-optimized LC/MS/MS System

Optimized System

8.0

7.0

7.5

8.0

min

min