[ CARE AND USE MANUAL ]

3

Figure 4: Waters Ferrule in a Parker Style Endfitting

The best way to fix the problem is to cut the tubing, replace the
ferrule and make a new connection.

1. Alternatively, replace the conventional compression screw fitting with

an all-in-one PEEK

fitting (Waters part number PSL613315) that

allows resetting of the ferrule depth. Another approach is to use a
Keystone, Inc. SLIPFREE

®

connector to always ensure the correct fit.

The fingertight SLIPFREE connectors automatically adjust to fit all
compression screw type fittings without the use of tools (Figure 5).


Figure 5: Single and Double SLIPFREE Connectors




SLIPFREE Connector Features:

• Tubing pushed into endfitting, thereby guaranteeing

a void-free connection

• Connector(s) come(s) installed on tubing

• Various tubing IDs and lengths available

• Fingertight to 10,000 psi – never needs wrenches

• Readjusts to all column endfittings

• Compatible with all commercially available endfittings

• Unique design separates tube-holding function from sealing function

Table 1: Waters Part Numbers for SLIPFREE Connectors

2. Flush column with 100% organic mobile phase (acetonitrile with

TEAA or methanol for TEA-HFIP ion-pairing method) by setting
the pump flow rate to 0.1 mL/min and increase the flow rate to
0.5 mL/min over 5 minutes.

3. When the mobile phase is flowing freely from the column outlet,

stop the flow and attach the column outlet to the detector. This
prevents entry of air into the detection system and gives more
rapid baseline equilibration.

4. Gradually increase the flow rate as described in step 3.

5. Once a steady backpressure and baseline at 260nm have been

achieved, proceed to the next section.

b. Column Equilibration

XBridge OST C

18

columns are shipped in 100% acetonitrile. It is

important to ensure mobile phase compatibility before changing
to a different mobile phase system. Equilibrate the column with a
minimum of 10 column volumes of the mobile phase to be used for
the oligonucleotide separation.

Note: When mobile phase additives are present in low concentrations
(e.g., TEAA or TEA-HFIP ion-pairing reagents), 100 to 200 column
volumes may be required for complete XBridge OST C

18

column

equilibration.

Gap

SLIPFREE Type

Tubing Internal Diameter

Tubing Length

0.005”

0.010”

0.020”

Single 6 cm

PSL 618000

PSL 618006

PSL 618012

Single 10 cm

PSL 618002

PSL 618008

PSL 618014

Single 20 cm

PSL 618004

PSL 618010

PSL 618016

Double 6 cm

PSL 618001

PSL 618007

PSL 618013

Double 10 cm

PSL 618003

PSL 618009

PSL 618015

Double 20 cm

PSL 618005

PSL 618001

PSL 618017