Waters ACQUITY APC Columns User Manual
Page 7
Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com
c. Column Use Information
The eCord Chip provides the customer with column use data, column dimensions and serial number. The overall
column usage information includes the total number of samples, total number of injections, total sample sets,
date of first injection, date of last injection, maximum pressure and temperature. The information also details
the column history by sample set including date started, sample set name, user name, system name, number
of injections in the sample set, number of samples in the sample set, maximum pressure and temperature in
the sample set, and if the column met basic system suitability requirements. Up to 50 sample sets can be
stored on the eCord Chip. In addition, the eCord provides two-way communications between the eCord Chip and
Empower
®
Software.
Waters, The Science of What’s Possible, Empower, UPLC, Advanced Polymer Chromatography, and ACQUITY are registered
trademarks of Waters Corporation. ACQUITY APC, APC, and eCord are trademarks of Waters Corporation.
©2014 Waters Corporation. Produced in the U.S.A. July 2014 720004667EN Rev. B TC-PDF