beautypg.com

B. calibration, C. useable flow-rate ranges – Waters ACQUITY APC Columns User Manual

Page 4

background image

4

ACQUITY APC Columns

b. Calibration

Whenever replacing a single column of a complete column bank, generate a new calibration curve to
ensure the reproducibility of the application. Figure 3 shows typical calibration curves for each column.
The calibration curves were obtained with polystyrene standards. Figure 4 shows the expected linear
range based on the calibration data.

100

1,000

10,000

100,000

1,000,000

10,000,000

0.50

1.00

1.50

2.00

Po

ly

st

yrene

Mole

cula

r We

ig

ht

(D

a)

Elution volume (mL)

ACQUITY APC
XT 450

ACQUITY APC
XT 200

ACQUITY APC
XT 900

ACQUITY APC
XT 125

ACQUITY APC
XT 45

Figure 3. Calibration Curves for ACQUITY APC XT and ACQUITY APC AQ Columns (4.6 x 150 mm) Using Polystyrene Standards.

10

10

1

10

2

10

3

10

4

10

5

10

6

10

7

Effective Molecular Weight Range

ACQUITY APC (AQ/XT) 45

ACQUITY APC (AQ/XT) 125

ACQUITY APC (AQ/XT) 200

ACQUITY APC (AQ/XT) 450

ACQUITY APC (XT) 900

Figure 4. Effective Molecular Weight Range for ACQUITY APC Columns.

c. Useable Flow-Rate Ranges

Excessive mobile-flow rates can create backpressure that can damage the column packing material.
For any given particle size, as the pore volume increases, the base particle loses mechanical strength
due to the reduction in particle wall thickness surrounding the open pore. Table 3 shows the maximum
flow rate recommendation for a single column for each pore size, independent of column length. When
connecting multiple columns in series, the column that specifies the lowest flow rate will dictate the
maximum flow rate for the column bank. Higher viscosity solvents will limit the flow rate due to flow
restriction. Table 4 provides viscosity data for common solvents.