beautypg.com

Care and use manual – Waters 30 mm ID Preparative Guard Cartridges and Guard Holders User Manual

Page 2

background image

2

[ CARE AND USE MANUAL ]

30 mm ID Preparative Guard Cartridges and Guard Holders

III. TIGHTENING PROCEDURE

1. After inserting guard cartridge, tighten holder by hand

until O-rings make contact with the guard cartridge.

2. While holding the 2-1/16” flats, use a 3/4” wrench to

tighten holder cap until O-rings are fully seated, between
5-9 tick marks.

IV. START-UP RECOMMENDATIONS

a. Background (Avoid Pressure Pikes)

It is possible that a SUDDEN high flow rate start at 50-60 mL/min
may contribute to initial disruption of the prep guard cartridge
bed leading to the peak splitting and shortened lifetimes.

It is recommended to start the flow rate up MORE SLOWLY
(over 5 or 10 minutes) to avoid this issue.

b. Start-Up

1. Purge the pumping system of any buffer-containing mobile

phases and connect the inlet end of the column to the injector
outlet.

2. Flush column with 100% organic mobile phase (methanol or

acetonitrile) by setting the pump flow rate to 0.1 mL/min and
increase the flow rate to 50-60 mL/min over 5 minutes.

3. When the mobile phase is flowing freely from the column

outlet, stop the flow and attach the column outlet to the
detector. This prevents entry of air into the detection system
and gives more rapid baseline equilibration.

4. Gradually increase the flow rate as described in step 2.

5. Once a steady backpressure and baseline have been achieved,

equilibrate the column with a minimum of 10-column volumes
of the mobile phase.

6. For added assurance, perform an efficiency test on the prep

guard cartridge before using it. Waters recommends using
a suitable solute mixture, as found in the “Performance Test
Chromatogram”, to analyze the prep guard cartridge upon
receipt. Determine the number of theoretical plates (N) and
use this value for periodic comparisons.

V. INSTALLING AND REPLACING THE GUARD
CARTRIDGE AND CARTRIDGE HOLDER

Install the cartridge holder onto the HPLC system in front of the
column using the rigid connector. The tubing in the rigid connector is
0.020” or 0.5 mm ID. For all other connections, use the same tubing.

Installing and replacing the guard cartridge:

1. Remove the cartridge holder from the system.

2. Unscrew the holder cap from the holder body.

3. Remove the old guard cartridge, if present.

4. Visually inspect the O-rings for damage or debris.

Replace, if necessary (see Section V).

5. Install the new guard cartridge. The direction of the flow

arrow on the guard cartridge and holder body should point
towards the detector.

6. Re-assemble the holder and tighten according

to instructions above.

7. Attach inlet side of assembled cartridge holder to the tubing

from HPLC injector.

8. Flush with 10 volumes of mobile phase at a low flow rate.

9. Attach outlet side of the cartridge holder to the column

inlet using rigid connector.

10. Check for leaks.

Figure 1. Preparative Guard Cartridge.

Guard Holder Body

Tick Marks

3/4” Wrench

2-1/16” Flats

Guard Cartridge

Guard Holder Cap