Maintenance and Servicing

8.3

Adjusting door

A well-closing door is indispensable for incubators.

On Memmert appliances, the tight closing of

the door is optimally guaranteed by a chamber

seal and a door seal (see also page 10 ). In

permanent operation, it is possible that the

flexible seal material will begin to sag. To ensure

that the door closes exactly despite this, an

adjustment may be necessary ( Fig. 38 ).
The top section (1) of the door hinge can be

moved slightly in the direction of the arrow after

the two screws (2) at the top and bottom of the

door have been undone. Afterwards, tighten the

screws down again.
Adjusting the door:
1. Undo the maggot screw (5) (with a jolting

motion, as it is fixed with locking paint).

2. Adjust the door by turning the eccentric (3)

with a screwdriver.

3. Apply locking paint to the maggot screw

and re-tighten it.

The locking plate ( Fig. 39 ) can also be adjusted

in the direction of the arrows after undoing the

screw (6).
1. Undo screw (6).
2. Move locking plate (7) in direction of arrow.
3. Tighten the screws again.

8.4

Repairs and Service

Warning!

After removing covers, live

parts may be exposed. You

may receive an electric shock

if you touch these parts. Disconnect the mains plug before remov-

ing any covers. Any work inside the unit may only be performed by

qualified electricians.
Repairs and service work are described in a separate service manual for the INCO

incubator.

2

1

3

4
5

Fig. 38

Adjusting door

1 Upper section of the door hinge

2 Mounting

bolts

3 Eccentric

tappet

4 Eccentric

socket

5 Headless

screw

7

6

Fig. 39

Adjusting locking plate

6 Screw

7 Locking

plate