beautypg.com

Memmert INCO 246/246 med CO2 Incubators (Generation 2003) User Manual

Page 34

background image

34

Operation and control

Weekday
Adjustment range:

Monday to Sunday

Mo

Tu

We

Th

Fr

Sa

Su

Day groups
Adjustment range:

Working days Mo-Fr

Weekend Sat-Sun

Mo

Tu

We

Th

Fr

Sa

Su

Mo

Tu

We

Th

Fr

Sa

Su

No switch on time:

----

Appliance not switched on on this day

on

off

Switch on time (on)
Adjustment range:

00:00 to 23:59 hours

on

off

h

Switch off time (off)
One minute beyond the switch on time up to

24:00

on

off

h

By turning further to the right, parameters (temperature, humidity setpoints etc.) can be

selected as in the normal operating mode.

If no settings (temperature setpoint etc.) are made for the ON phase, the controller takes

over the values from the normal operating mode.

For reasons of safety, you should always check that only one switch one time is programmed

in the desired time blocks and days. By turning further to the right, parameters (temperature

setpoint etc.) can be selected as in the normal operating mode.
If the controller is in standby mode or the week time switch is in the ON phase, the tempera-

ture setpoint can be directly accessed by briefly pressing the SET key. By turning the control to

the right, you are returned to temperature monitoring, humidity and CO

2

setting. By turning

to the left, you come back to the settings for the individual time blocks.