Advanced functions

7.3.3 Automatic temperature monitor ( ASF)

ASF is a monitoring device that automatically follows the set temperature setpoint within an

adjustable tolerance band ( Fig. 25 ).
The ASF is activated – if switched on – automatically if the actual temperature value reaches

50 % of the set tolerance band of the setpoint (in the example: 37 °C – 1.5 °C) reached for the

first time (section A). The activation of the ASF is shown by the brightly lit

AUTO

icon.

When the temperature moves outside the set tolerance band around the setpoint (in the

example in Fig. 25 : 37 °C ± 3 °C) – e. g. if the door is opened during operation (section B of

illustration) – the alarm is set off. This is shown by the flashing

AUTO

and icons.

If the acoustic alarm is switched on in the SETUP, the ASF alarm is additionally signalled by an

intermittent tone. If the SET key is pressed, the horn can be temporarily switched off until the

next alarm event occurs.
The ASF alarm goes off automatically as soon as 50 % of the set tolerance band of the set-

point (in the example: 37 °C ± 1.5 °C) are reached again (section C).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in the

example: The setpoint is changed from 37 °C auf 30 °C, section D), until it has reached the

tolerance range of the new temperature setpoint (section E).

40 °C

34 °C

ASF active

40 °C

34 °C

ASF alarm

ASF active

ASF active

33 °C

27 °C

°C

37 °C

t

A

B

C

D

E

AUTO

AUTO

AUTO

Fig. 25

Schematic diagram of how the ASF temperature monitoring functions

Switching on the automatic temperature monitor:

Select the AUTO icon with the push-turn control.
Hold down the SET key and set to on with the push-turn control.

MIN

MAX

AUTO

°C

Switching off the automatic temperature monitor:

Select the AUTO icon with the push-turn control.
Hold down the SET key and set to off with the push-turn control.

The tolerance band for the ASF can be set in the SETUP in the

range 0.5 ... 5 °C (see page 46 ).