4. Setting the humidity setpoint:
Turn the push-turn control to the right until the humid-

ity display flashes. Hold down the SET key and set the

desired humidity setpoint of 96.0 %rh with the push-turn

control. After releasing the SET key, the humidity setpoint

briefly flashes. The current humidity value appears on

the display and the controller begins to move to the set

value.

The humidification process is indicated by the

icon

(only for models with humidity module).

rh

%

5.

Setting the CO

2

setpoint

Turn the push-turn control to the right until the CO

2

dis-

play flashes. Hold down the SET key and set the desired

CO

2

setpoint of 5.0 % with the push-turn control. The

appliance flashes briefly, showing the CO

2

setpoint. The

current CO

2

actual value appears on the display and the

controller begins to move to the set CO

2

setpoint.

Depending on the bottle used, fumigation is dis-

played by the

IN 1

or

IN 2

icon.

CO

2

6. Adjust the O

2

setpoint

(only for models with O

2

module)

Turn the push-turn control to the left until the O

2

display

flashes. Hold down the set-key and set the desired O

2

setpoint of 3.0 % with the push-turn control. Release

the set-key. The appliance flashes briefly, showing the O

2

setpoint. Then, the current O

2

actual value appears on

the display and the controller begins to move to the set

O

2

value.

on

off

Mo

Tu

We

Th

Fr

Sa

Su

o

2

= 3 . 0 %

The incubator is now running in permanent operation with the set values.

5.10.3

Week time switch

In this operating mode, the appliance switches on and

off automatically at the times programmed.
During the OFF phase of the week time switch, the

appliance is in standby mode. The heating and cooling functions, along with the CO

2

and

humidity supply are switched off here and the controller display shows the time, dimmed.
The sequence of the week time switch repeats itself each week.
In total, a maximum of 9 time blocks can be programmed, consisting of the switching on and

switching off times.

PRINT