Advanced functions

Setting:

Overtemperature limit:
up to max. 10 °C above nominal temperature

(for details of the nominal temperature, see nameplate)
Setting:
1. Select the

MAX

icon with the push-turn control.

2. Hold down the SET key and set the desired temperature limit

with the push-turn control (e.g. 38.5 °C).

°C

MIN

AUTO

MAX

Undertemperature limit:
Adjustment range: 10 °C below the minimum temperature of the

appliance up to 10 °C above nominal temperature of the appli-

ance (for details of the nominal temperature, see nameplate).

The lower alarm limit value cannot be set higher than the top

one. If no undertemperature limit is required, set the lowest

temperature.

°C

MIN

AUTO

MAX