Operation and control

5.10.1

Normal mode

The appliance runs in this operating mode in permanent operation. The desired setpoints for

operating the chamber can be selected. The settings have an immediate effect on the func-

tions of the appliance.
1. Load incubator (see page 27 ).
2. Switch on appliance. To do this, press the push-turn control on the operating panel so that

it comes out of the appliance (see Fig. 16 on page 28 ).

3. Select the normal operating mode

with the push-turn control:

PRINT

SETUP

4. As described above, set the individual parameters with the push-turn control and the set

key:

Temperature setpoint

Adjustment range: 20°C to 50°C

°C

Temperature monitoring
Adjustment range:

MIN MAX AUTO

(see also page 47 )

°C

MIN

AUTO

MAX

Humidity setpoint

Adjustment range:

For incubators with basic

fittings: 88 to 97 %rh

For models with humidity

module: 40 to 97 %rh

rh

%

CO

2

setpoint

Adjustment range: 0 bis 10 %

(or models with CO

2

or premium

modules 0 to 20 0 to 20 %)

CO

2

O

2

setpoint

(only for models with O

2

mod-

ule)
Adjustment range:

Off, 1 to 20 %

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

o

2

= 3 . 4 %