Operation and control
5.10.1
Normal mode
The appliance runs in this operating mode in permanent operation. The desired setpoints for
operating the chamber can be selected. The settings have an immediate effect on the func-
tions of the appliance.
1. Load incubator (see page 27 ).
2. Switch on appliance. To do this, press the push-turn control on the operating panel so that
it comes out of the appliance (see Fig. 16 on page 28 ).
3. Select the normal operating mode
with the push-turn control:
SETUP
4. As described above, set the individual parameters with the push-turn control and the set
key:
Temperature setpoint
Adjustment range: 20°C to 50°C
°C
Temperature monitoring
Adjustment range:
MIN MAX AUTO
(see also page 47 )
°C
MIN
AUTO
MAX
Humidity setpoint
Adjustment range:
►
For incubators with basic
fittings: 88 to 97 %rh
►
For models with humidity
module: 40 to 97 %rh
rh
%
CO
2
setpoint
Adjustment range: 0 bis 10 %
(or models with CO
2
or premium
modules 0 to 20 0 to 20 %)
CO
2
O
2
setpoint
(only for models with O
2
mod-
ule)
Adjustment range:
Off, 1 to 20 %
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
o
2
= 3 . 4 %