Delivery, Transport and Setting Up

3.3.4 Storage after delivery

If the incubator is initially to be stored after delivery: Read the storage conditions from page

65 .

3.4

Setup

The incubator can be placed either on the ground or on a table (work surface). When doing

this, ensure that the appliance is positioned exactly horizontally

.

The installation site must be

level and able to reliably carry the weight of the incubator (see page 18 ). Do not place the ap-

pliance on an inflammable surface.
A power connection must be available at the installation site in accordance with the connec-

tion data on the nameplate (see page 18 ).
The distance between the wall and the rear of the chamber must be at least 15 cm. The clear-

ance from the ceiling must not be less than 20 cm and the side clearance from the wall not

less than 8 cm ( Fig. 7 ). Sufficient air circulation in the vicinity of the chamber must be guaran-

teed at all times.

•FP

•FP

•FP

•FP

Fig. 7

Minimum clearance from walls and ceiling