Design and Function

Digitalised electronic CO

2

control with automatic zero position, NDIR measuring system

with self-diagnosis system and acoustic error display, air pressure compensation

Language settings (see page 46 )

Alphanumeric text display

Integrated week time switch with grup function (e.g. each working day) (see page 33 )

Retracting push-turn control for simple operation of appliance (see page 28 )

Two separate Pt100 temperature sensors DIN Class A in a 4-wire circuit for control and

monitoring

Digital monitoring control for overtemperature, undertemperature and automatic

setpoint following (ASF) (see page 50 )

Mechanical temperature limiter (TB protection class 1, see page 47 )

Monitoring relay to switch off heater in case of error

Optical alarm display

Acoustic signal messages if temperature or CO

2

limits are crossed, when door is open and

if gas bottle is empty (see page 40 )

Calibration of temperature, humidity, CO

2

and O

2

possible on the device without a sepa-

rate PC (see from page 53 )

2.3.2 Optionally available additional modules
Comfort module:

Two gas connections with quick release connectors (see page 13 )

Automatic switch-over of gas bottles

Hygiene module

Electro-polished, seamless laser-welded chamber

Communication module

Logging option of temperature, CO2 and relative humidity via computer/laptop

Internal log memory with 1024 kB as ring memory for all setpoint and actual values, errors

and settings in real time and with date, logging approx. 3 months at 1 minute storage

interval (see page 62 )

Parallel printer port (PCL3-compatible) for printing out log data (see page 13 and page 45 )

optionally USB, Ethernet, RS-232 or RS-485 interface (see from page 58 )

CO

2

module

Extended adjustment range from 0 to 20 %

3-point calibration (5 %, 10 %, 15 % CO

2

)

O

2

module

Control of oxygen concentration through the introduction of nitrogen (N

2

); adjustment

range 1 % to 20 % O

2

(not in combination with Comfort or Premium module)

Premium module

Includes comfort, hygiene, communication and CO

2

modules

Humidity module

Active microprocessor humidification and dehumidification control (40-97 % rh)