Advanced functions

2. With a calibrated reference instrument and at a set setpoint temperature of 35 °C in nor-

mal operation, an actual temperature of 34.6 °C is measured.

3. Set compensation correction value for CAL.2 in the SETUP to - 0.4 °C:

h

°C

°C

°C

4. The reference instrument after the calibration procedure should now display 35 °C.
5. With CAL.1, another compensation temperature below CAL.2 can be programmed in the

same way, and with CAL.3, one lying above this.

If all compensation correction values are set to 0.0 °C, the factory calibration settings are

restored.

7.6.2 Humidity calibration

The incubator can be calibrated for the individual customer by means of two balance points

at 20 and at 90 % relative humidity. For each selected balance point a positive or negative

compensation correction value can be set between –5 % and +5 %.

rh 20

+3,0 %

rh 90

-2,0 %

60%

40%

0%

20%

80%

100%

Default calibration

Fig. 29

Humidity calibration (example)

Setting:

1. Set the desired humidity balance point in the SETUP (see page 46 ) and set accompanying

compensation correction value to 0.0 °C.

2. With a reference instrument, measure the deviation in the stationary state in the selected

humidity balance point.

3. Setting the compensation correction value in the SETUP. If the measured reference humid-

ity is too low, the compensation correction value must be set with a negative sign.

4. Perform a control measurement with the reference instrument.
The procedure can be performed with humidity balance points of 20% rh and 90% rh.