Memmert INCO 246/246 med CO2 Incubators (Generation 2003) User Manual

Page 42

background image

42

Warning messages and malfunctions

Error

Possible causes

Remedy

Nothing shows on the

display although the

incubator is switched

on.

Power supply

interrupted

Check power supply and fuse/safety

switch.

Appliance fuse or

miniature fuse or

controller faulty

Contact the customer service and read

the service manual.

Appliance error

Mainboard faulty

Appliance cannot be

operated

Incubator locked with

user-ID card

Unlock incubator with user-ID card (see

page 52 ).

Push/turn control

faulty

Contact the customer service and read

the service manual.

No CO

2

display shown

in the CO

2

module

Switched mode power

supply SP 200 faulty

Contact the customer service and read

the service manual.

icon flashes ...

Temperature fuse

(TWW, ASF) has re-

sponded

Increase temperature difference

between monitoring and working

temperature (see page 47 ).

Replace Pt100 temperature sensor

of monitoring controller if necessary

(see Service manual)

... and rh empty

Water supply tank

empty

Set humidity setpoint to OFF, fill up

distilled water, then reset humidity

setpoint back to desired value.

... and rh empty

Humidity setpoint

exceeded

Open door for 30 sec. and wait to see if

subsequently the controller steadily ad-

justs to the setpoint. If the error occurs

again, Contact the customer service.

... and co2 over

CO

2

setpoint is ex-

ceeded

Open door for 30 sec. and wait to see if

subsequently the controller steadily ad-

justs to the setpoint. If the error occurs

again, Contact the customer service.

... and error Autozero

Autozero pump

faulty

CO

2

controller

faulty

Contact the customer service and read

the service manual.

Error display (E...) in

display

Appliance/system error Contact the customer service and read

the service manual.