Page 30: 10 setting the operating mode

background image

30

Operation and control

The control returns automatically to the main menu if the push-turn key or set key is

not operated for approx. 30 seconds.

Setting the temperature (Quick adjustment):

1. Hold down the SET key and set the desired temperature setpoint with the push-turn

control.

2. Release the SET key
The appliance flashes briefly, showing the temperature setpoint. Then the current temperature

appears on the display and the controller begins to move to the set temperature.

5.9

Operating modes

INCO incubators can be operated in three ways ( Fig. 19 ).

Normal mode: The incubator runs in permanent operation at the temperature, humidity

and CO

2

values set on the operating panel. Operation in this mode is described from page

31 .

Week time switch: The incubator runs at the set values only at certain times. Operation in

this mode is described from page 33 .

Interface mode with PC/laptop (for models with communication or premium modules, see

from page 58 )

In addition, basic appliance settings can be made (SETUP, see page 45 ) and printouts can be

made if the appliance is equipped with the communication or premium module (PRINT, see

page 45 ).

PRINT

SETUP

Normal

operation

(see Page 31 )

Week time

switch

(see Page 33 )

Printer

(see Page 45 )

Basic appliance

settings

(see Page 45 )

Fig. 19

Operating modes

5.10 Setting the operating mode

1. Hold SET key down for approx. three seconds, the selected operating mode

then begins to flash.

2. Select the desired operating mode (normal mode, week time switch, pro-

gramming mode, printer or basic appliance settings/setup by turning control

with SET key held down.

3. Release the SET key, and the selected operating mode is saved.

set

set

set