Warning messages and malfunctions

Error

Possible causes

Remedy

Error message CONF
(is displayed for only 10

sec. after switching on)

Checksum error (error

when saving setpoint

values)

The error can be rectified just by the

controller after a setpoint parameter

has again been saved. If the error

should continue occurring, or cannot be

rectified: Contact the customer service

and replace controller (see service

manual)

Chamber fan without

function

Miniature fuse or

mains adapter faulty

Contact the customer service and read

the service manual.

Heating icon not on

Ambient tempera-

ture too high

Setp the appliance up in a cooler room,

minimum setpoint temperature = am-

bient temp. + 8 °C

Temperature in

appliance higher

than set setpoint

temperature

Wait until appliance has cooled down.

icon is permanently

lit up

Temperature fuse has

responded

Switch off appliance and leave to

cool down

Check temperature limiter

The appliance is only operational again

after it has cooled down and the error

has been eliminated.

CAL O

2

ERROR

Fault in oxygen calibra-

tion

Open the door for one minute and

start calibration again. If the message

appears gain: Contact the customer

service.

ERROR O

2

Sensor

Oxygen sensor faulty

Contact the customer service