Design and Function

2.2

Function

The air in the incubator is heated up by an all-round heater with a large surface ( Fig. 3 , No. 1).

The CO

2

and/or N

2

gas is introduced into the working chamber via a sterile filter. Because CO

2

has a much higher specific weight than air, the gas is let into the working chamber above the

interior fan (2). The turbulence-free interior ventilation (3) ensures a uniform distribution of

the gases, creating a homogenous atmosphere.
In the basic version, humidification takes place by means of water trays (4). To avoid uncon-

trolled condensation, dosed fresh air is piped into the interior. If the appliance is equipped

with a humidity module, humidification takes place via a hot-air generator, which allows

water to evaporate at a set rate. The sterile hot air is let into the interior above the fan and is

mixed with the air current (5).

1

1

1

1

1

1

4

4

2

5

3

1

Fig. 3

How INCO incubators work

2.3

Optional extras

2.3.1 Basic equipment

Electronic fuzzy-supported PID process controller with pulse width modulation and per-

manent performance adjustments and time-saving self-diagnosis system to quickly locate

errors (see page 41 )

All-round heater with larger surface area, with additional thermal conduction layer (see

Fig. 3 )

Capacitive humidity sensor

Humidity control (fresh air via sterile filter) ensures that setpoint humidity is quickly

reached and guarantees short recovery times, while avoiding condensation formation

Homogenous atmosphere and temperature distribution through encapsulated, turbu-

lence-free ventilation system

STERICard for fully automatic sterilisation process control for hot air sterilisation of appli-

ance, including sensors and fan rotor (see page 51 )