Warning messages and malfunctions

6.

Warning messages and malfunctions

6.1

Warning messages

An intermittent tone is also set off by the warning messages. This can be temporarily

switched off by pressing the SET key.

DOOR OPEN if the door of

appliance is opened for

longer than 3 minutes.

Remedy: Close the door

Error in the temperature control system (see also Chapter “ Temperature monitoring ”

on page 47 ):

Tb active - temperature

limiter triggered if the

temperature limiter

responds

Remedy: Switch off the appliance

and leave to cool down. If the

error occurs again after switch-

ing the

appliance back on: Switch off

the appliance and Contact the

customer service

hi-alarm - overtempera-

ture alarm limit max ex-

ceeded if overtemperature

protection responds

Remedy: Check the setting of

the MAX temperature monitor-

ing (see Chapter “ Temperature

monitoring ” on page 47 .

lo-alarm - undertem-

perature alarm limit min

crossed if undertempera-

ture

protection responds

Remedy: Check the setting of

the MIN temperature monitor-

ing (see Chapter “ Temperature

monitoring ” on page 47 .

asf alarm - temperature

outside tolerance band if

automatic monitoring

function responds

Remedy: Check the setting of the

ASF temperature monitoring (see

Chapter “ Automatic temperature

monitor (ASF) ” on page 50 .

Error in humidity system (only for appliances with humidity module)

if the water supply is

defective

Remedy: Check that the water

supply hose is properly con-

nected. Fill the water supply tank

with distilled water if it is empty.