beautypg.com

C a l o, O f f, 4 oxygen calibration – Memmert INCO 246/246 med CO2 Incubators (Generation 2003) User Manual

Page 57: Advanced functions example: co, Content of 10.0 %, an actual co, Calibration settings are restored, Only for models with o2 module) in the o, Off is displayed. release the set-key. 2. set co

background image

57

Advanced functions

Example: CO

2

deviation in chamber load at 10 % should be corrected:

1. Set the CO

2

balance point in the SETUP to CO2 10 and set the accompanying compensation

correction value to 0.0 %:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

IN 1

IN 2

OUT

MIN

AUTO

MAX

2. With a calibrated reference instrument and at a set CO

2

content of 10.0 %, an actual CO

2

content of 11.5 % is measured.

3. Set the compensation correction value in the SETUP for CO2 10 to 1.5 %:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

IN 1

IN 2

OUT

MIN

AUTO

MAX

4. The reference instrument should display 10 % after the calibration procedure.
With CO2 5 and CO2 15, further calibrations can be programmed at 5 % and 15 % (15 % only

for models with CO

2

or premium modules).

If all compensation correction values are set to 0.0 %, the factory CO

2

calibration settings

are restored.

7.6.4 Oxygen calibration

(only for models with O2 module)
In the O

2

calibration, the oxygen proportion in the incubator at normal atmosphere is deter-

mined.
1. Switch off the O

2

module, if active. To do this, turn the

push-turn control to the left until the O

2

display flashes.

Hold down the set-key and turn the push-turn control

until O

2

off is displayed. Release the set-key.

2. Set CO

2

setpoint value to 0% (see page 33 ).

3. Open the oven door and inner glass door, in case of

four-part door, remove glass panes. Keep the doors

open for at least one minute so that a normal atmos-

phere can form in the chamber. Close the doors again.

4. In the setup, select the item O

2

calibration with the

push-turn control (cal 02 off). Hold down the set key

and set to cal 02 on with the push-turn control. Release

the set key and leave the setup via EXIT.

Now, the oxygen content in the incubator is measured

for about one minute. In the display, cal o2 active can be

seen. Do not switch off the oven while this is taking

place.

5. When the calibration is finished, the oxygen content

determined is shown in the display:

on

off

Mo

Tu

We

Th

Fr

Sa

Su

o

2

o f f

off

C A L o

2

o n

C A L o

2

a c t i v e

o

2

= 2 0 . 7 %