Warning messages and malfunctions

6.3

Power failure

In case of a power failure, the incubator operates as follows:

In normal and week time switch operating modes

After the power supply has been restored, operation is continued with the parameters set.

The time and the duration of the power failure is documented in the log memory.

For remote operation

If there is a power failure in remote operation, the appliance starts in the normal operating

mode for safety reasons and all setpoint values are changed to safe default values (see table).

The programme can only be continued from the computer. The time and duration of the

power failure are documented in the log memory.

Parameters

Default value

Temperature 20

°C

CO

2

0

%

rh off
O

2

(only for models with O

2

module)

OFF