Memmert INCO 246/246 med CO2 Incubators (Generation 2003) User Manual
Page 46

46
Advanced functions
By turning the push-turn control and holding down the SET key, the following parameters can
be selected and altered one after another, as described in Chapter “ Basic operation “ on page
29 :
Clock time in 24 hr. format
Conversion to summer time is not automatic,
but must be done manually.
Date
The controller contains a calendar which automatically
accounts for the different lengths of months, and for leap
years.
Weekday
Tu
Year
Adjustment range: from 2000 to 2100
Acoustic signal at programme end
ENDSOUND
Setting: OFF or ON
Acoustic Signal for alarm, e.g. over/undertemperature
ALARM so
Setting: OFF or ON
Communication address
Adjustment range: 0 to 15 (see Chapter
“ Communication interfaces “ on page 58 )
Tolerance band ASF
Adjustment range: 0.5 bis 5°C (see page 50 )
Language
Setting: GERMAN, English, franCAIS, ESPANOL and italIANO
Corrective values (CAL 1-3, rh20, RH90) customer-side calibra-
tion of temperature and humidity (see Chapter “ Calibration “
on page 53 )