Delivery, Transport and Setting Up
3.4.1 Base (accessory)
The incubator can be placed on a base ( Fig. 8 ).
3.4.2 Stacking frame (accessory)
Two appliances of the same model size can be placed on top of one another. To do this, foot
alignment provisions must be attached to the lower oven ( Fig. 9 ):
1. Remove the housing cover from the lower oven.
2. Insert drilling template (supplied with the foot alignments) into the overturned lid.
3. Mark drilling points and drill with a 4.2 mm diameter drill bit.
4. Screw the foot alignment provisions to the top of the lid with the screws and nuts sup-
plied.
5. Put the covers back on.
Fig. 8
Base
Fig. 9
Assembly of the foot alignment
provisions when two incubators are
placed on top of one another