Advanced functions

Error rectification after the TB has been triggered:

1. Switch off appliance and leave to cool down
2. Contact the customer service and have the error rectified (e.g. replace temperature sen-

sor).

The appliance is only operational again after it has cooled down and the error has been elimi-

nated.

7.3.2 Electronic temperature monitoring ( TWW)

Temperature monitoring can be adjusted independently of the operating modes.
The manually set monitoring temperature

MIN

and

MAX

the overtemperature control is

monitored by an adjustable over/undertemperature controller (TWW) protection class 3.3 acc.

to DIN 12880.
If the manually set monitoring temperature

MAX

is exceeded, the TWW takes over tempera-

ture control and begins to regulate the monitoring temperature (Fig. 21). The alarm icon

flashes as a warning .

Setting MAX

Set
temperature

t

°C

Controller error

Emergency operation

Fig. 24

Schematic diagram of how the TWW temperature monitoring functions

The monitoring temperature must always be set sufficiently high above the maximum

working temperature.

If the acoustic alarm is switched on in the SETUP, the TWW alarm is additionally signalled by

an intermittent tone. If the SET key is pressed, the horn can be temporarily switched off until

the next alarm event occurs.