4 sterilisation chipcard – Memmert INCO 246/246 med CO2 Incubators (Generation 2003) User Manual

Page 51

background image

51

Advanced functions

7.4

Sterilisation chipcard

For decontamination, the incubator is

equipped with a STERICard as standard

( Fig. 26 ). The STERICard starts an

automatic and predetermined sterilisa-

tion sequence.

Warning!

The incubator becomes

very hot inside during

the sterilisation proce-

dure. Do no open the

door.

Caution:

The sterilisation programme is not

intended for sterilising the chamber

load, but only for sterilising the inte-

rior of the appliance. The incubator

is not a steriliser with respect to the

Law on Medical Devices.

Make sure that the chamber is empty before the sterilisation procedure is started. Sliding

shelves and removed water trays may also be sterilised.
To activate,the STERICard, insert it into the chip card reader (see page 10 ) on the operating

panel of the appliance and select the Start icon

with the push-turn control while the SET

key is held down.
As soon as the sterilisation procedure has been enabled, the automatic sterilisation procedure

begins:
Ramp 1: Heating up to 160 °C
Ramp 2: Hold time 4 hours (after setpoint temperature has been reached)
Ramp 3: Cooling down to 70 °C
Ramp 4 Wait time of 10 minutes

The automatic sterilisation programme cannot be changed.

After the sterilisation has been completed, STERILISATION OK is shown in the controller display.

For documentation purposes, the sterilisation procedure is logged on the STERICard and can

be read out with a computer/laptop.

Sterilisation

Process Control

Card

Sterilisation Parameter:

160 °C / 4:00 h

Unit-ID:

________________________

Fig. 26

STERICard