Cleaning and maintenance – Leica Biosystems ASP200 S User Manual
Page 80
80
Instructions for Use V 1.4, RevE – 03/2013
7. Cleaning and Maintenance
Cleaning the surface of the infiltration module
Important in order to ensure that the seals always seal properly.
• Remove both covers for cleaning.
• First, clean the stainless steel surface using the plastic spatulas, then
wipe to remove all dirt particles around the retort and the paraffin res-
ervoirs.
Reagent and condensate bottle seal lubrication
• To ensure easy removal of the reagent and condensate bottles, lubricate
the O-ring seals on the plug-in nozzles with the O-ring lubricant supplied.
This procedure is particularly important with bottles containing
chloroform. O-rings that have not been lubricated swell when ex-
posed to chloroform. The reagent containers can then become very
difficult to remove.
Checking the retort lid seal
• Regularly check the retort lid seal for damage. If the seal is damaged, it
must replaced without delay.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)