57

Leica ASP200 S

5. Operation

Remote draining
• On the screen, select the reagent container

(No. 3) and the receptacle (External).

Then press STATION DRAIN.
• Check that no reagent has leaked.

Remote filling
• On the screen, select an empty reagent

container (No. 3) and the external receptacle

(External).

Then press the STATION REFILL button.
• The remote filling should be completed in less

than 170 seconds.

• Check that no reagent has leaked.

Select station

Press button

Filling/draining reagents
From the FAVORITES screen, press MENU. In

the MENU FUNCTIONS window, tap the SMART

SCREEN button.
The SMART SCREEN (

Fig. 55

) pops up. The smart

screen is the initial screen for manual operation

of the instrument.

Fig. 55

Fig. 56

Status bar,

displayed in supervisor mode.

Status bar
The following values are displayed.
• Pressure and temperature of the retort

• Wax bath temperature

• Value of the level sensor of the retort

• Retort lid - closed or open

The following steps should only be

carried out by trained laboratory

personnel, experienced in handling

reagents.