Operation – Leica Biosystems ASP200 S User Manual
Page 57
57
Leica ASP200 S
5. Operation
Remote draining
• On the screen, select the reagent container
(No. 3) and the receptacle (External).
Then press STATION DRAIN.
• Check that no reagent has leaked.
Remote filling
• On the screen, select an empty reagent
container (No. 3) and the external receptacle
(External).
Then press the STATION REFILL button.
• The remote filling should be completed in less
than 170 seconds.
• Check that no reagent has leaked.
Select station
Press button
Filling/draining reagents
From the FAVORITES screen, press MENU. In
the MENU FUNCTIONS window, tap the SMART
SCREEN button.
The SMART SCREEN (
Fig. 55
) pops up. The smart
screen is the initial screen for manual operation
of the instrument.
Fig. 55
→
Fig. 56
Status bar,
displayed in supervisor mode.
Status bar
The following values are displayed.
• Pressure and temperature of the retort
• Wax bath temperature
• Value of the level sensor of the retort
• Retort lid - closed or open
The following steps should only be
carried out by trained laboratory
personnel, experienced in handling
reagents.