beautypg.com

3 smart clean, Cleaning and maintenance – Leica Biosystems ASP200 S User Manual

Page 76

background image

76

Instructions for Use V 1.4, RevE – 03/2013

7. Cleaning and maintenance

7.1.3 Smart clean

The smart cleaning prepares the ASP200 S for transport. During a smart

clean, a series of air cleaning steps will be applied which will remove any

excess liquid (condensate) from all internal tubing.
Before starting a smart cleaning, a complete retort cleaning program must

be carried out in the ASP200 S.
To start a smart clean, press the SMART CLEAN button on the MORE MENU

FUNCTIONS screen and press OK to confirm.
The smart clean itself takes approx. 15 minutes and consists of 3 phases:

Phase 1:
Prior to starting the smart clean make sure all reagent containers are cor-

rectly inserted and connected.

All internal tubing is cleaned with compressed air to force any remaining

reagents back into the reagent containers.

Phase 2:
Before phase 2 starts, the operator is prompted to remove all reagent con-

tainers from the instrument.

All internal tubing is again cleaned with compressed air.

Phase 3:
When phase 3 starts, the operator is prompted to place an empty container

in station 11.

The tubing is once again cleaned with pressurized air; any remaining re-

agent residues are forced back into station 11.

Empty and clean all reagent containers thoroughly before reinsert-

ing them into the instrument to make sure there will be no reagent

spills from the bottles that could cause damage during transport.