Instrument components and specifications – Leica Biosystems ASP200 S User Manual
Page 20
20
Instructions for Use V 1.4, RevE – 03/2013
3. Instrument Components and Specifications
Retort
Capacity:
max. 200 cassettes
Reagent volume:
3.5 l with volume replacement body
Temperature (paraffin):
40 to 65
°C
Temperature (processing reagents):
Ambient temperature or 35 to 55
°C
Temperature (cleaning reagents):
50 to 65 °C
Temperature accuracy:
+ 1 K
Filling time:
approx. 90 sec
Drain time:
approx. 80, 120, 140 sec (selectable)
Impregnation vacuum:
-70 kPa (g)
Infiltration pressure:
35 kPa (g)
Filling vacuum:
-70 kPa (g)
Draining pressure:
35 kPa (g)
General
Reagent container:
10
Cleaning solution bottles:
3
Maximum bottle volume:
4.0 l
Printer:
Optional Accessories
Pretest check:
ON/OFF
Fluid level sensor:
ON/OFF
Recirculation (pump in/out):
ON/OFF
(a) Time before 1st cycle:
16 min
(b) Time between cycles:
20 min
System setup
Password status:
Supervisor/Operator
Type of password:
alphanumeric, freely selectable
Reagent management system (RMS):
ON/OFF
Software interlock:
ON/OFF
3.4 Technical data (continued)