3 troubleshooting, Troubleshooting – Leica Biosystems ASP200 S User Manual
Page 71
71
Leica ASP200 S
6. Troubleshooting
6.3 Troubleshooting
→
Use the SMART SCREEN functions to individually test operations
such as fill, drain, pressurizing and evacuation operations.
If you cannot identify the problem by carrying out the above listed
steps, call Leica Technical Service.
If a problem occurs in the Leica ASP200 S, take the following steps to diag-
nose the problem:
• Check for error messages indicating the nature of the problem.
• If an error message is displayed, press the Help button on the message
for any information relating to the cause.
• Check the Run Log for any messages that indicate the nature of the fail-
ure. If the failure occurred during a program run, check the Run Log for
an indication of whether the failure occurred during the fill, processing
or drain cycles, and the reagent container in use at the time.
Power failures
If there is evidence of power to the instrument:
• Verify that the power plug is plugged into the socket and that the socket
is switched ON.
• Check whether the bridge cable is plugged in properly.
• Verify that the power switches on the rear of the instrument (adjacent to
the power lead) and on the side of the instrument are both switched ON.
• Messages indicating that there has been a partial power failure indicate
that power has been lost to the heaters, but not to the main control elec-
tronics. Carry out the above steps to identify the nature of the problem.