7. Cleaning and Maintenance
Replacing the active carbon filter
• The life of the active carbon filter will depend
on the reagent types used and the frequency
of vacuum cycles.
• The filter (13) should be replaced at least every
3 months. (
Fig. 86
)
After replacing the filter, open the SMART
MONITOR menu to reset the filter age to "0".
For additional information, refer to
Chap. 5.5
• We recommend a smart clean prior to chang-
ing the filter.
Fig. 86
Draining the condensate container
• Remove, empty, and reinsert the blue conden-
sate container (
Fig. 85
).
Fig. 85
Filter label
14
13
Dispose of waste solvents with care
according to local regulations and the
waste management policy of the com-
pany or institution.
When inserting the new filter, make
sure that it is inserted with the cor-
rect side up, as specified on the label
on the front side of the filter.The arrow
must point upwards.