5. Operation
5.2.2 Replacing the paraffin
Draining the paraffin
• Connect the paraffin drain hose (36) to the
spout (17) of the paraffin drain at the front of
the instrument (
Fig. 57
) and into a catch tank.
Fig. 57
• Select the paraffin reservoir to be emptied and
the receptacle (External,
Fig. 58
).
Then, press STATION DRAIN....
• This is followed by multiple security prompts
that you have to confirm individually in order
to continue.
17
36
From FAVORITES, call up the SMART SCREEN
screen. (
see previous chapter
)
→
+
→
Fig. 58
When connecting the hose, be sure to
press it onto the O-rings of the drain
opening as far as it will go.
The paraffin drain hose must be
securely hooked into the external re-
ceptacle and remain there through the
entire drainage process.
When drainage is finished, the hose is
cleaned automatically using air.
Do not remove the hose from the exter-
nal receptacle until this cleaning step
is complete.