1 cycle for changing reagents, Instrument components and specifications – Leica Biosystems ASP200 S User Manual
Page 23
23
Leica ASP200 S
3.6.1 Cycle for changing reagents
Reagent
Week 1
Week 2
Week 3
Week 4
Formalin
At least 3 times a
week
At least 3 times a
week
At least 3 times a
week
At least 3 times a
week
Formalin
At least 3 times a
week
At least 3 times a
week
At least 3 times a
week
At least 3 times a
week
70 %
Daily
Daily
Daily
Daily
80 %
Once a week
Once a week
Once a week
Once a week
95 %
Once a week
---
Once a week
---
95 %
---
Once a week
---
Once a week
100 %
Once a week
---
Once a week
---
100 %
---
Once a week
---
Once a week
Xylene
Once a week
---
Once a week
---
Xylene
---
Once a week
---
Once a week
Paraffin 1
Run paraffin clean-
ing program* daily.
Replace the paraf-
fin after 6 cleaning
cycles.
Run the paraffin
cleaning program*
once a week.
Run the paraffin
cleaning program*
twice a week.
Run paraffin clean-
ing program* daily.
Replace the paraf-
fin after 6 cleaning
cycles.
Paraffin 2
Run the paraffin
cleaning program*
twice a week.
Run paraffin clean-
ing program* daily.
Replace the paraf-
fin after 6 cleaning
cycles.
Run the paraffin
cleaning program*
once a week.
Run the paraffin
cleaning program*
twice a week.
Paraffin 3
Run the paraffin
cleaning program*
once a week.
Run the paraffin
cleaning program*
twice a week.
Run paraffin clean-
ing program* daily.
Replace the paraf-
fin after 6 cleaning
cycles.
Run the paraffin
cleaning program*
once a week.
Cleaning xylene
Once a week
Once a week
Once a week
Once a week
Cleaning alcohol
Once a week
Once a week
Once a week
Once a week
Cleaning Water
Once a week
Once a week
Once a week
Once a week
* Paraffin cleaning program,
see Chap. 7.1.2
3. Instrument Components and Specifications