3. Instrument Components and Specifications

3.2 Specific instrument options

• Optionally usable Reagent Management System (RMS), displays service life and usage fre-

quency of the individual reagents and enables automatic definition of the reagent sequence

– there is no more need to rearrange reagent containers. If one or more reagents within a

sequence are replaced, the RMS automatically uses the reagents in the correct sequence,

organized in ascending order of cleanliness.

• Liquid movement ("wave motion") during the process for better and continuous mixture of

the reagents.

• System for non-contact filling/draining of reagents – drains and fills the reagent container

using a hose connected to the infiltration module without the operator being exposed to

reagents in the process.

• Non-contact wax bath drainage.
• Optical level meter.
• Active paraffin cleaning program – removes solvent residue from the paraffin, lengthening

its service life.

• Magnetic stirrer – for gentle circulation of the reagents, thus ensuring a uniform reagent

temperature.

• Programmable end time for infiltration programs.
• 3-step drainage of the retort (adjustable) for reducing reagent displacement.
• Infiltration process at ambient pressure, or with pressure, vacuum or a combination of both.
• Four programmable cleaning programs. The cleaning programs automatically omit all steps

that are not necessary to complete the cleaning procedure.