Instrument setup – Leica Biosystems ASP200 S User Manual
Page 33
33
Leica ASP200 S
SYSTEM DIAGNOSTICS menu
If TOUCH HERE ... is touched during
initialization (
Fig. 14
) and the pass-
word for supervisor mode is en-
tered, the SYSTEM DIAGNOSTICS
menu (
Fig. 16
) opens. This menu
provides access to basic instru-
ment settings.
Notice!
Only experienced operators may
adjust the settings, since incorrect
use of the functions can result in
serious malfunctions.
4. Instrument Setup
Fig. 16
Fig. 17
The following functions can be selected:
1 - Access for service technicians only.
2 - Displays the INSTALLATION menu.
3 - Aborts the current program.
4 - Aborts the current program and deletes the
current allocation of reagents to bottles and
retort.
5 - Deletes all reagents and resets programs and
instrument status (all lists will be empty).
6 - Saves the current instrument status.
7 - Displays the results log.
8 - Restarts the instrument.
To exit this menu, the device must be restarted.
press RESTART APPLICATION.
To reinitialize the instrument, press YES to con-
firm the query
"ARE YOU SURE YOU WANT TO ..."
(Fig. 17).
3
4
2
1
5
8
7
6
Queries such as this are always made
before important steps that are irre-
versible.
This allows the operator to undo
changes caused by accidental key
strokes.
• The initialization process starts again with the
screen shown in
Fig. 14
.