Operation – Leica Biosystems ASP200 S User Manual
Page 38

38
Instructions for Use V 1.4, RevE – 03/2013
PROGRAM OPTIONS
Select the options with which programs are executed here. Press a button
to change the associated value.
• STATION SEQUENCE: By Age or Sequential
By Age - RMS is activated, reagents will automatically be used in
order of increasing cleanliness.
Sequential - reagents will be used in sequential order of stations.
• WAX BATH ORDER:
Auto or 1. ; 2. ; 3.
Can only be activated if STATION SEQUENCE is set to "Sequential" -
otherwise the order of baths will be managed by the RMS.
• PROMPT FOR NUM BLOCKS:
Enabled or Disabled
Enabled- When starting a program, the actual number of blocks must
be entered. This is recommended with the RMS being en-
abled.
Disabled - Number can be entered, but does not have to be.
DATE / TIME
Be sure to verify that the date and time entered do actually correspond
to the local time/date, as this will ensure that all programs are carried out
correctly.
Press the buttons to access the entry windows where settings or correc-
tions can be made, if necessary.
5. Operation
Date and time setup screen
Fig. 27
Fig. 25
Fig. 26