Entering / modifying reagent thresholds
If warning thresholds are required for a certain reagent, enter them as fol-
lows:
• Highlight the reagent to be modified, either by pressing the reagent name
or using the UP/DOWN buttons.
• Press the header of the parameter to be changed; the corresponding
entry screen appears.
• Enter a new threshold value or - if no warning is desired - press CLEAR
to remove the threshold altogether.
• Press OK to confirm.
5. Operation
Changeable parameters
Fig. 42
Fig. 43
Allocation of reagents to the correct reagent group is the basis for compatibility monitoring.
Allocation to the wrong reagent group can lead to reagent cross contamination.
The respective threshold value applies for all reagent stations con-
taining the same reagent.
If a reagent is already used in a program, it can neither be renamed
nor can it be allocated to another reagent group!
The corresponding symbols will be disabled (i.e. they will not be
surrounded by a blue border).
If a reagent is renamed, all stations and programs linked to that re-
agent need to be reedited as well!
Changing reagent names or reagent groups
• Highlight the reagent the name or group of which you wish to change.
• Press the corresponding button symbol in the headline.
• In the entry window (or via the keyboard), enter the new reagent group
allocation / the new reagent name.
• Press OK to save the new reagent group / reagent name.