Initial operation – Leica Biosystems ASP200 S User Manual
Page 28
28
Instructions for Use V 1.4, RevE – 03/2013
Fig. 8
Inserting the volume replacement body
26
25
27
25
19
Important!
The instrument must be operated only
when the volume replacement body is
in the retort (even during cleaning pro-
grams).
Otherwise, malfunctions may cause
damage to the instrument.
• Take the volume replacement body (
Fig. 8,
25)
at the metal clip (26) and insert it into the retort
(19) such that the bore (27) is above the posi-
tion of the magnetic stirrer.
4. Initial Operation
4.3 Installing accessories (continued)
Fig. 7
Magnetic stirrer
• Unpack the magnetic stirrer (
Fig. 7,
30) and
insert it into the retort.
• Important!
The magnetic stirrer must be inserted such
that the outer curvature (32) of the two stir-
rer blades (31) faces downwards!
(
Fig. 7
)
Do not insert them the other way around, as
then there will be no effective stirring.
32
31
30
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)