beautypg.com

2 reagent handling, 1 draining/filling reagents (other than paraffin), Reagent handling – Leica Biosystems ASP200 S User Manual

Page 56: Operation

background image

56

Instructions for Use V 1.4, RevE – 03/2013

5. Operation

5.2 Reagent handling
5.2.1 Draining/filling reagents
(other than paraffin)

1. Manually changing the reagents
• Remove the appropriate reagent container from the reagent cabinet,

remove the bottle screw cap.

• Drain/refill the reagent by pouring to/from a bulk container. Use the

included funnel for clean filling.

• On the REAGENT STATUS screen, verify correct allocation and set the

reagent container as full.

2. Use the "Remote Fill / Drain" features

• Connect the hose for remote filling/draining

as shown in

Fig. 6

to the connection (7).

Fig. 54

7

While performing a reagent fill/drain, ensure that the remote fill/

drain hose is securely placed in the remote container and is not re-

moved from the container until the operation is fully completed

Finally, pressurized air is used to clear the hose after each fill /

drain so that no dirt remains.

The hose should therefore not be removed from the bulk container

until this cleaning step is complete.

Important!

Press the connection of the hose firmly

into the holder until you can clearly

hear it click.

• Hook the other end of the hose into the bulk

container that you are emptying or filling.