5.5 System monitor

The SYSTEM MONITOR function provides important infor-

mation about the status of the Leica ASP200 S.

After replacing the filter, press the CLEAR button to set the age to "0".
ROTARY VALVE RETRIES:
Provides information about the status of the rotary valve. If the number

specified there exceeds 50, maintenance is recommended.

PUMP AGE:
Indication of the operating hours of the pressure/

vacuum pump. The configured warning value

determines the triggering of a warning message.

The pump must be maintained by Leica Service

after approx. 1000 operating hours.

AGE OF THE CARBON FILTER:
Displays the operating hours of the active carbon

filter. The configured warning value determines

the triggering of a warning message. The carbon

filter must be replaced after approx. 3 months.

Fig. 78

FILL / DRAIN:

The table shows the measured fill-

ing and drain times of all stations

(in seconds) as average values.

Short-term = the last 5 operations
Long-term = the last 20 operations
The values are usually different, but should not

have any extremes. If one of the values is signifi-

cantly above 250, contact Leica Service.

There could be a blockage, which in a worst-

case scenario could cause the instrument to

malfunction.