5 system monitor, System monitor, Operation – Leica Biosystems ASP200 S User Manual
Page 68
68
Instructions for Use V 1.4, RevE – 03/2013
5. Operation
5.5 System monitor
The SYSTEM MONITOR function provides important infor-
mation about the status of the Leica ASP200 S.
After replacing the filter, press the CLEAR button to set the age to "0".
ROTARY VALVE RETRIES:
Provides information about the status of the rotary valve. If the number
specified there exceeds 50, maintenance is recommended.
→
→
PUMP AGE:
Indication of the operating hours of the pressure/
vacuum pump. The configured warning value
determines the triggering of a warning message.
The pump must be maintained by Leica Service
after approx. 1000 operating hours.
AGE OF THE CARBON FILTER:
Displays the operating hours of the active carbon
filter. The configured warning value determines
the triggering of a warning message. The carbon
filter must be replaced after approx. 3 months.
Fig. 78
FILL / DRAIN:
The table shows the measured fill-
ing and drain times of all stations
(in seconds) as average values.
Short-term = the last 5 operations
Long-term = the last 20 operations
The values are usually different, but should not
have any extremes. If one of the values is signifi-
cantly above 250, contact Leica Service.
There could be a blockage, which in a worst-
case scenario could cause the instrument to
malfunction.
Fig. 79