2 access levels, Operation – Leica Biosystems ASP200 S User Manual
Page 40
40
Instructions for Use V 1.4, RevE – 03/2013
SECURITY
Access rights to the instrument are managed here as
user profiles.
5. Operation
SUPERVISOR MODE: Enabled or Disabled
Enabled- There are two different user profiles.
A password is required for supervisor level access to the
in- strument. Enabling SUPERVISOR MODE already requires
entering the password.
5.1.2 Access levels
The Leica ASP200 S may be configured to allow two levels of user access.
Operator access level:
• Operators may run programs and view results. On this level, the OPERA-
TOR symbol is displayed in the upper right corner of the touchscreen; all
enabled buttons are surrounded by a black border.
Supervisor access level:
• Supervisors may perform all Operator functions, and additionally create
programs and perform the instrument setup functions.
In supervisor mode, the SMART SCREEN menu displays a status bar with
information about the retort and the wax baths (see
Fig. 55
)
• To access Supervisor level, press SUPERVISOR, enter the required
password and confirm. After you enter the password, the SUPERVISOR
symbol appears in place of OPERATOR, the outlining of all active buttons
changes from black to blue, the SUPERVISOR button is replaced by the
OPERATOR button.
"Operator" symbol
"Supervisor" symbol
→
Disabled -
Default state of the instrument. All instrument and software
functions of the ASP200 S are fully accessible to all staff.
Fig. 31
Supervisor mode is disabled in factory condition.