SECURITY

Access rights to the instrument are managed here as

user profiles.

5. Operation

SUPERVISOR MODE: Enabled or Disabled

Enabled- There are two different user profiles.

A password is required for supervisor level access to the

in- strument. Enabling SUPERVISOR MODE already requires

entering the password.

5.1.2 Access levels

The Leica ASP200 S may be configured to allow two levels of user access.

Operator access level:
• Operators may run programs and view results. On this level, the OPERA-

TOR symbol is displayed in the upper right corner of the touchscreen; all

enabled buttons are surrounded by a black border.

Supervisor access level:
• Supervisors may perform all Operator functions, and additionally create

programs and perform the instrument setup functions.

In supervisor mode, the SMART SCREEN menu displays a status bar with

information about the retort and the wax baths (see

Fig. 55

)

• To access Supervisor level, press SUPERVISOR, enter the required

password and confirm. After you enter the password, the SUPERVISOR

symbol appears in place of OPERATOR, the outlining of all active buttons

changes from black to blue, the SUPERVISOR button is replaced by the

OPERATOR button.

"Operator" symbol

"Supervisor" symbol

Disabled -

Default state of the instrument. All instrument and software

functions of the ASP200 S are fully accessible to all staff.

Fig. 31

Supervisor mode is disabled in factory condition.