4 technical data, Technical data, Instrument components and specifications – Leica Biosystems ASP200 S User Manual
Page 19

19
Leica ASP200 S
3. Instrument Components and Specifications
Nominal supply voltages:
Two factory-preset voltages (not user-adjustable):
100 to 120 V
or
230 to 240 V
Nominal frequency:
50 to 60 Hz
Main fuses:
2 melting fuses, 20 x 5 mm, UL-approved
• for 100 to 120 V F 10 A 250 VAC
• for 230 to 240 V F 5 A 250 VAC
Nominal power:
1000 VA
Dimensions, (L x W x H), in mm:
595 x 680 x 1325
Empty weight, unpacked:
approx. 160 kg
Weight, packed:
220 kg
Operating temperature range:
15 °C to 35 °C
Relative humidity:
10 % to 80 % non-condensing
IEC 1010 classification:
Protection class 1
Pollution degree 2
Overvoltage installation category II:
• 800 V impulse (120 V systems)
• 1500 V impulse (240 V systems)
Altitude:
2500 m maximum
Local/remote alarm relay:
30 V DC, maximum 2 A
2 terminals:
Each with isolated switching contact (operable both
as normally-open and normally-closed circuit)
Paraffin reservoirs
Number of containers:
3
Capacity:
4.3 l per container
Melting time:
approx. 10 h
Temperature:
40 to 65 °C
Temperature accuracy:
+ 1 K
3.4 Technical Data