Operation – Leica Biosystems ASP200 S User Manual
Page 39
39
Leica ASP200 S
INSTRUMENT
Press PARAFFIN BATH SET TEMP. to open the input screen (
Fig. 30
) for the
wax bath temperature.
Set the temperature according to the value required by the paraffin used.
Select the highest permissible temperature for the paraffin to ensure that a
minimum loss of temperature occurs when filling the retort.
DISPLAY/BUZZER:
• The number next to DISPLAY OFF indicates the remaining time (in min)
until the screen saver (after the last user action) is activated.
• The BUZZER VOLUME can be set between 1 and 10.
5. Operation
Paraffin temperature and
run options screen
Fig. 30
The RUN OPTIONS
displays the options with
which a program starts.
The options set apply for
ALL programs!
A printer option will be dis-
played only if a printer has
been configured for the
instrument.
The individual options are
described in
Chap. 5.3.1
.
Fig. 29
Fig. 28
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)