INFICON CMS5000 Monitoring System User Manual

Page 24

background image

3 - 4

IP

N 07

4-

50

8-

P1

C

CMS5000 Operating Manual

the filled sampling vessel. The CMS5000 weighs 55.1 lbs. (25 kg). All plumbing to
and from the sampling vessel should be supported independently of the analytical
module, as the system design will not support any additional weight.

For air monitoring, the CMS5000 should be mounted in an optimal location for
collecting air samples of interest. Care should be exercised to avoid locations
affected by air currents from heating or cooling systems, doors, windows or
structural openings.

To ensure proper operation when utilizing the CMS5000 for continuous water
monitoring, it is necessary to determine optimal mounting height and unit
placement to accommodate both the water intake as well as allow for clearance for
the sample collection tube when attaching and removing the sampling vessel.
Allow a minimum of a 15” (38.1 cm) clearance below the mounted sampling vessel
for vessel removal. Verify that the CMS5000 is level when mounted and that the
slope is less than five percent. See

Figure 3-2

.

Figure 3-2 Dimensions for Mounting the CMS5000