CMS5000 Operating Manual

12

Each individual compound will need to be manually added to the library
template. To select a peak, refer to the instructions in

section 12.8, How to

Access the Scan Cursor, on page 12-25

. See

Figure 16-10

.

Figure 16-10 Selecting a Peak

13

Once a peak is selected, right-click and select Grab Peak for Template to add
a compound to the library template. See

Figure 16-11

.

Figure 16-11 Grab Peak for Template

14

The following table will be displayed in the center of the screen. Enter the name
of the analyte in the highlighted field. See

Figure 16-12

.

NOTE: The following screenshots depict the calibration of a water method.

However, the same steps are used when creating air and check
standard methods.

Figure 16-12 Entering Analyte Name