Series connected eritech – ERICO 3000 ERITECH SYSTEM User Manual
Page 54

INSTALLATION, OPERATION AND MAINTENANCE MANUAL
52
www.erico.com
2 x FRP Mast
(2 m or 4.6 m)
ERITECH
®
DYNASPHERE
Guy Kit
Aluminum Mast
U Bolts
2 x ERITECH
®
ERICORE
Cable
Guy Ring
1 x TERMDUALCOUP
The twin ERITECH
®
ERICORE conductors
are run in a parallel manner down to 
ground (see Figure 36). The usual 
conductive saddle spacing applies.
At the point of connection to the 
grounding system, each ERITECH
®
ERICORE cable is fitted with an ERITECH 
ERICORE lower termination kit, and 
attached to the grounding system
Series connected ERITECH
®
ERICORE cable sections
As an alternative to running twin 
ERITECH ERICORE conductors, a single 
ERITECH ERICORE cable can be run, but 
it is broken into several sections, none 
of which exceeds the cable breakdown 
voltage. This is done by periodically 
connecting the cable to a conductive 
structural point (see Figure 37).
IMPORTANT – To maintain the 
necessary clearance between 
the ERITECH
®
DYNASPHERE air
terminal and the top of the mast, the 
FRP mast must protrude at least 
1400 mm above the top of the 
aluminum mast. The following shows 
the required positioning dimensions for 
the two available FRP mast sizes.
2 meter FRP Mast
Overlap on aluminum mast = 600 mm
Protrusion above aluminum mast = 1400 mm
4.6 meter FRP Mast
Overlap on aluminum mast = 1.5 m
Protrusion above aluminum mast = 3.1 m
The standard guy ring part may be fitted to 
the Dual Coupling Termination and used 
with the usual guying kit.
Figure 35: Typical mast arrangement using the twin ERITECH
®
ERICORE arrangement.
