Transport, storage and disposal
11.1
Transport
Use the original packaging for transport.
11.2
Inserting transport locks
Air temperature
Relative humidity
Atmospheric pressure
General transport
-25 °C – 60 °C
10 % – 95 %
30 kPa – 106 kPa
Air freight
-40 °C – 55 °C
10 % – 95 %
30 kPa – 106 kPa
NOTICE! Material damage due to missing transport securing device.
The device may be damaged if it is transported without the transport securing device.
Insert the transport securing device before you transport the device.
1. Make sure the device is switched off.
2. Close the plate compartment flaps.
3. Remove the plate compartment by hand.
4. Place the transport lock on the plate
compartment.
5. Use the two inner screws to screw the transport
lock and plate compartment together.
6. Manually retract the plate compartment.
7. Screw down the transport lock (1) using the two
external screws (2).
8. Close the flaps for the plate compartment and
filter slides.
2
2
1