Technical data
Eppendorf
®
PlateReader AF2200
English (EN)
61
9
Technical data
9.1
Power supply
9.2
Weight/dimensions
9.3
Ambient conditions
9.4
Measuring methods
• Upper fluorescence intensity
• Lower fluorescence intensity
• Absorbance
9.4.1
Notes on measuring methods
• Microplates with 6 to 384 wells are suitable for all measuring methods if they comply with the following
standards:
– ANSI/SLAS 1-2004
– ANSI/SLAS 2-2004
– ANSI/SLAS 3-2004
– ANSI/SLAS 4-2004
• The device can conduct kinetic measurements.
• The measurement can be limited to one part of the microplate.
Mains connection
100 – 120 V/220 – 240 V, 50 – 60 Hz
Power consumption
150 VA
Overvoltage category
ll
PlateReader AF2200
16 kg
Dimensions
Width
462 mm (18.19")
Height
282 mm (11.1")
Depth
458 mm (18.03")
Ambient temperature during operation
15 °C to 30 °C (59 °F to 86 °F)
Ambient temperature, not during operation
-20 °C to 60 °C (-4 °F to 140 °F)
Relative humidity during operation
< 90 %, non-condensing
Degree of pollution
2