beautypg.com

5 inserting the filter slide, 1 information on optical filters, 1 fluorescence filters – Eppendorf AF2200 Plate Reader User Manual

Page 29: 2 absorbance filters, Inserting the filter slide 5.1, Information on optical filters 5.1.1, Fluorescence filters, Absorbance filters

background image

Inserting the filter slide

Eppendorf

®

PlateReader AF2200

English (EN)

29

5

Inserting the filter slide

5.1

Information on optical filters

5.1.1

Fluorescence filters

The optical filters (band-pass filter) were especially developed for fluorescence measurements. The spectral
permeability and bandwidth of the fluorescence filters guarantee maximum sensitivity.

5.1.2

Absorbance filters

Band-pass filters which are used for absorbance measurements in microplate readers have a bandwidth of
5 to 10 nm. Therefore, fluorescence filters should not be used for absorbance measurements because the
bandwidth is normally larger than 10 nm.

During the measurement of dyes with narrow peaks, an incorrect filter can lead to a bright value error or
low absorbance values.

Eppendorf offers two predefined filter slides. They are delivered preconfigured and can be
used immediately for measurements.
The filters are permanently installed on these predefined filter slides. They cannot be removed
or exchanged.